Light BYTES: March 2019- Learn how the SPECTRA III light engine facilitates faster Image Acquisition.

SPECTRA III light engine versus SPECTRA X light engine:
Increased Brightness For Faster Image Acquisition

The metric most important when evaluating light intensities required for widefield fluorescence microscopy is the irradiance (or power density) at the sample plane, expressed in mW/mm2.

Our data table shows irradiance generated by SPECTRA III® and SPECTRA X® light engines in the four principle excitation bands used in multicolor fluorescence microscopy. Irradiance required for widefield fluorescence microscopy is typically on the order of 10–100 mW/mm2. Clearly, such levels are provided by the SPECTRA X light engine; prompting the question, why would the higher irradiance levels provided by the SPECTRA III be useful?

Higher irradiance allows exposure times to be shortened while maintaining the number of fluorescence photons detected. Shorter exposure times provide increased temporal resolution in time-lapse image sequences and reduce the time required to acquire multicolor z-stacks or slide scans.

Lumencor’s SPECTRA III light engine provides the best of spectral breadth, brightness, and speed for even the most demanding imaging and high throughput applications.

Measurement Conditions:

  • Power measurements made with Coherent FieldMax II–TO power meter with PM3 thermopile detector. Nikon Ti microscope with Semrock FF409/493/573/652-Di02 quad dichroic.
  • Excitation filters (CWL/FWHM, nm): DAPI = 390/22; FITC = 475/28; TRITC = 555/28; Cy5 = 635/22.
  • Light engine output coupled to microscope using Lumencor 3 mm diameter liquid light guide (PN 10-10589) and collimating adapter (PN 82-10121).
Download the PDF of Light BYTES: March 2019